MPI - Thermal - ThermalAir TA-3000A ThermalAir Series Temperature Forcing System
Flexible Temperature Testing & Thermal Cycling
The ThermalAir temperature products are a series of high capacity thermal air stream systems that are used for temperature testing, fast thermal cycling and device temperature characterization of components, hybrids, modules, PCBs, and other electronic and non-electronic assemblies at precise temperatures from -65°C to +225°C
System Performance
Airflow: Continuous Air Flow up to 16 SCFM @-65°C / +225°C
-40°C to +125°C / +125°C to -40°C in Less Than 10 sec
Temperature Control: Environment Air or External DUT sensing directly at the DUT case to ±1.0°C
Temperature control Sensor Ports: Type T or Type K thermocouple (2x Rear), and 100 ohm RTD
Temperature set, display and resolution: ±0.1°C
Temperature Accuracy: ± 1.0°C (calibrated system)
Remote interface ports: Four: IEEE-488, RS232, SOT/EOT/SFF, 4 USB-Type A, 1 USB-Type B, LAN, PLUS Auto Start Test & End of Test for automatic temperature cycling Hot-Cold-Amb.
All this makes for simple control and service ability.
The ThermalAir temperature products are a series of high capacity thermal air stream systems that are used for temperature testing, fast thermal cycling and device temperature characterization of components, hybrids, modules, PCBs, and other electronic and non-electronic assemblies at precise temperatures from -65°C to +225°C
System Performance
Airflow: Continuous Air Flow up to 16 SCFM @-65°C / +225°C
-40°C to +125°C / +125°C to -40°C in Less Than 10 sec
Temperature Control: Environment Air or External DUT sensing directly at the DUT case to ±1.0°C
Temperature control Sensor Ports: Type T or Type K thermocouple (2x Rear), and 100 ohm RTD
Temperature set, display and resolution: ±0.1°C
Temperature Accuracy: ± 1.0°C (calibrated system)
Remote interface ports: Four: IEEE-488, RS232, SOT/EOT/SFF, 4 USB-Type A, 1 USB-Type B, LAN, PLUS Auto Start Test & End of Test for automatic temperature cycling Hot-Cold-Amb.
All this makes for simple control and service ability.